2019

Electrochemiluminescent biosensor with DNA link for selective detection of human IgG based on steric hindrance.

Talanta Liu Q, Xie XL, Mao CJ, Chen JS, Niu HL, Song JM
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A highly selective DNA-based electrochemiluminescence (ECL) based biosensor is described for the detection of human IgG. It is exploiting the effect of steric hindrance that affects the strength of the ECL signal in the presence of IgG. Digoxin-linked signaling DNA was specifically bound to IgG, and this causes steric hindrance which limits the ability of DNA to hybridize with capturing DNA attached to a gold electrode. Europium (II) doped CdSe quantum dots were covalently linked to the DNA in order to generate the ECL signal. Using this steric hindrance hybridization method, the ECL signal of the biosensor were proportional to the concentration of IgG with a wide linear range and a 14 pM detection limit. Conceivably, the method can be expanded to the detection of a wide range of proteins for which homologous recognition elements are available.

关键词