2019

Dual rapid test for HIV and syphilis: A laboratory evaluation of the diagnostic accuracy of the Standard Q HIV/Syphilis Combo Test.

Diagnostic microbiology and infectious disease Bristow CC, Rivera SKV, Ramos Cordova LB, Palacios LJQ, Konda KA, Klausner JD
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We evaluated the Standard Q HIV/Syphilis Combo Test (SD Biosensor, South Korea), a dual rapid test using stored sera (N = 400) in a laboratory setting in Lima, Peru. The sensitivity and specificity for HIV antibody detection was 100.0% (95% CI: 98.2-100.0%) and 99.5% (95% CI: 97.2-100.0%), respectively. For treponemal antibody detection the sensitivity and specificity was 97.5% (95%CI:94.3-99.2%) and 100.0% (95%CI:98.2-100.0%), respectively.