2019

Probe-specific loop-mediated isothermal amplification magnetogenosensor assay for rapid and specific detection of pathogenic Leptospira.

Molecular and cellular probes Nurul Najian AB, Foo PC, Ismail N, Kim-Fatt L, Yean CY
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This study highlighted the performance of the developed integrated loop-mediated isothermal amplification (LAMP) coupled with a colorimetric DNA-based magnetogenosensor. The biosensor operates through a DNA hybridization system in which a specific designed probe captures the target LAMP amplicons. We demonstrated the magnetogenosensor assay by detecting pathogenic Leptospira, which causes leptospirosis. The color change of the assay from brown to blue indicated a positive result, whereas a negative result was indicated by the assay maintaining its brown color. The DNA biosensor was able to detect DNA at a concentration as low as 200 fg/μl, which is equivalent to 80 genomes/reaction. The specificity of the biosensor assay was 100% when it was evaluated with 172 bacterial strains. An integrated LAMP and probe-specific magnetogenosensor was successfully developed, promising simple and rapid visual detection in clinical diagnostics and service as a point-of-care device.