2019

On-line monitoring of heavy metals-related toxicity with a microbial fuel cell biosensor.

Biosensors & bioelectronics Adekunle A, Raghavan V, Tartakovsky B
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This study describes an environmental biosensor for real time toxicity monitoring, which exploits high sensitivity of a microbial fuel cell (MFC) to variations in concentrations of electron donors and acceptors. Fast biosensor response to changes in total heavy metal concentration of a mining rock drainage was observed in laboratory tests with acceptable repeatability and a coefficient of determination (R2) of 0.95-0.97. The biosensor response is attributed to interference of heavy metals with the activity of electroactive microorganisms. Biomolecular analysis of the anodic electroactive biofilms showed significant population differences between microbial populations of the biosensor exposed to heavy metals and a non-exposed (control) biosensor. Furthermore, the biosensor outputs were highly correlated (R2 = 0.92) with the results of the Microtox toxicity assay. The results of this study contribute towards the development of a simple MFC-based biosensor capable of detecting, in real-time, changes in environmental conditions and providing a tool for on-site toxicity monitoring.

关键词