2019

A novel ultrasensitive electrochemiluminescence biosensor for glutathione detection based on poly-L-lysine as co-reactant and graphene-based poly(luminol/aniline) as nanoprobes.

Biosensors & bioelectronics Wang C, Chen L, Wang P, Li M, Liu D
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work, an ultrasensitive electrochemiluminescence (ECL) biosensor was constructed using poly-L-lysine (PLL) as a novel co-reactant of luminol and poly(luminol/aniline) nanorods loaded reduced graphene oxide (PLA@rGO) as nanoprobe, which enable highly sensitivity detection of glutathione (GSH). To the best of our knowledge, it is the first time that PLL was used for the co-reactant of luminol. Notably, about a 5-fold enhancement was obtained compared with the individual PLA@rGO using GCE. Due to the remarkable quenching effect between the excited state of PLL and the reduced form of GSH in the ECL system of luminol/PLL, the ECL sensing platform exhibited wide linear ranges of 1.0 × 10-9-1.0 × 10-4 M and 1.0 × 10-4-1.0 × 10-2 M and a low detection limit of 7.7 × 10-10 M. Simultaneously, the biosensor was also successfully applied to detect GSH in human serum sample with high recoveries. Hence, this work would open a new platform for the wide application of PLL in immunoassay and various sensors.