2019

A voltammetric biosensor for mercury(II) using reduced graphene oxide@gold nanorods and thymine-Hg(II)-thymine interaction.

Mikrochimica acta Jin H, Zhang M, Wei M, Cheng JH
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The presented voltammetric mercury(II) sensor is based on the specific interaction between Hg(II) ion and thymine-thymine base pairs. Reduced graphene oxide is functionalized with gold nanorods and then loaded with thionine and streptavidin (RGO@AuNR-TH-SA). A T-rich thiolated DNA (S1) is firstly immobilized on a gold electrode. In the presence of Hg (II), the T-rich biotin-DNA (biotin-S2) binds to S1 via T-Hg(II)-T interaction. Then, the RGO@AuNR-TH-SA is linked to the gold electrode by specific binding between SA and biotin-S2. This produces an electrochemical signal (at -0.208 V vs. Ag/AgCl) of TH that depends on the concentration of Hg (II). The peak current increases linearly in the 1 to 200 nM Hg (II) concentration range, and the detection limit is 0.24 nM. The sensor is highly selective for Hg (II) over other environmentally relevant metal ions, even at concentration ratios of >1000. Graphical abstract Schematic representation of a voltammetric biosensor for mercury(II) using reduced graphene oxide@gold nanorods (RGO@AuNRs) and thymine-Hg(II)-thymine interaction. It is based on the fact that RGO@AuNR can strongly adsorb thionine (TH) and streptavidin to realize the signal amplification.

关键词