2019

Top-Down Strategy of Implantable Biosensor Using Adaptable, Porous Hollow Fibrous Membrane.

ACS sensors Zhou J, Ma Z, Hong X, Wu HM, Ma SY, Li Y, Chen DJ, Yu HY, Huang XJ
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Fabrication of an outer membrane is crucial for an implantable biosensor to enhance the long-term stability and accuracy of sensors. Herein, an adaptable, controllable, porous outer membrane for an implantable biosensor was fabricated using a "top-down" method, allowing maximum retention of enzyme activity and fine control over membrane microstructure. Polysulfone hollow fibrous membranes with different pore sizes and porosities were used as a base membrane. Chitosan (CH) and sodium alginate (SA) were self-assembled on the inner surface of PSfHM to construct a biocompatible and conductive interface between PSfHM and the electrode. In vitro and in vivo experiments were used to evaluate the performance of implantable glucose biosensors with PSfHM and CH/SA modified PSfHM (PSfHM-CH/SA). The glucose biosensor with PSfHM-CH/SA exhibited a more stable output current than bare sensors and a quick response time (<50 s). The glucose biosensor with PSfHM-CH/SA linear sensing range was between 0 and 22 mM ( R2 = 0.9905), and relative sensitivity remained at >87% within 7 days and >76% within 15 days. Furthermore, response currents recorded by implanted sensors closely followed the blood glucose trend from the tail vein blood during in vivo experiments.

关键词