2019

A New Analytical Method for Determination of Cathepsin L Based on the Surface Plasmon Resonance Imaging Biosensor.

International journal of molecular sciences Tokarzewicz A, Romanowicz L, Sankiewicz A, Hermanowicz A, Sobolewski K, Gorodkiewicz E
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

The purpose of this study was to develop a new method for a determination of the cathepsin L-biosensor based on the Surface Plasmon Resonance Imaging technique. The cathepsin L is an endopeptidase, which degrades proteins and plays an important role in various processes occurring in the human body. The detection technique, Surface Plasmon Resonance Imaging, is an optical, label-free technique, which can be used for quantitative determination of the different proteins. In order to bind the enzyme, the cathepsin L inhibitor-RKLLW-NH2 was used. The validation process showed that parameters: precision, accuracy, and selectivity of the method were acceptable. The analytically useful range of the standard curve was 0.50 ng/mL-15.00 ng/mL. The detection and quantification limit of method was 1.67 pg/mL and 5.07 pg/mL, respectively. The usefulness of the developed method was confirmed by the determination of the cathepsin L concentration in the blood plasma of some healthy persons and in the blood plasma of patients. The obtained results were compared with the results obtained by the ELISA. It was found that the correlation between these two methods was very strong, what suggest that the developed method can be used as the competitive method to the ELISA.

关键词