组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
A newly designed amperometric biosensor for the determination of ethanol through one-step electrochemical coating of (4,7-di(thiophen-2-yl)benzo[c][1,2,5]selenadiazole-co-1H-pyrrole-3-carboxylic acid) (TBeSe-co-P3CA) on a graphite electrode is presented. It was aimed to propose a newly synthesized copolymer with enhanced biosensing properties as a novel sensor for the quantification of ethanol. The conjugated copolymer (TBeSe-co-P3CA) was prepared through electrochemical polymerization by potential cycling. After polymer modification, alcohol oxidase (AOx) was immobilized on a modified electrode surface for ethanol sensing. In the analytical investigation, the calibration plot is linear above large concentration range (0.085 to 1.7 mM), where sensitivity is around 16.44 μA/mMcm2 with a very low detection limit (LOD) of 0.052 mM based on the signal-to-noise ratio in short response time. Moreover, interfering effect of some possible compounds were examined and the capability of the biosensor in estimating ethanol content in commercial alcoholic beverages was also demonstrated. The results showed satisfactory accuracy of the developed sensor and confirm the proposed sensor has a potential for ethanol quantification compared to the currently used techniques.