2019

Disposable electrochemical immunosensor for analysis of cystatin C, a CKD biomarker.

Talanta Lopes P, Costa-Rama E, Beirão I, Nouws HPA, Santos-Silva A, Delerue-Matos C
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Specific monitoring of cystatin C (CysC) levels in biological fluids is critical for diagnosis, treatment and mechanistic understanding of a spectrum of diseases, particularly chronic kidney disease (CKD). Despite evidences that CysC correlates with the high risk and/or progression of CKD, its use in clinical practice is still scarce. In this context, we report the development of a simple and sensitive immunosensor for the detection of CysC. The biosensor combines the technology of cost-effective screen-printed electrodes with the high specificity of a sandwich immunoassay. Optimized conditions showed that the sensor operates in a linear range between 10 and 100 ng mL-1, with a detection limit and a sensitivity of 6.0 ng mL-1 and 6.4 ± 0.3 μA ng mL-1 cm-2, respectively. Moreover, the sensor provided precise results (RSD ≤ 6.2%) and the quantification of CysC in CKD serum samples revealed to be in agreement with the values obtained by a particle-enhanced nephelometric immunoassay. In this light, the proposed immunosensor qualifies for clinical application, constituting a step forward in the development of fast, sensitive and cost-effective diagnostic tools that can improve the current medical care settings of CKD patients.