Fabrication of a novel biosensor for biosensing of bisphenol A and detection of its damage to DNA.

Talanta Jalalvand AR, Haseli A, Farzadfar F, Goicoechea HC
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work, a novel electrochemical biosensor has been fabricated based on step-by-step modification of a glassy carbon electrode (GCE) with methylene blue (MB)-DNA/multiwalled carbon nanotubes (MWCNTs)-chitosan (CS)/palladium nanoparticles (Pd NPs)/fullerene C60 (C60) for voltammetric and impedimetric detection of DNA damage induced by bisphenol A (BPA). Modifications applied to the GCE were characterized by cyclic voltammetry (CV), differential pulse voltammetry (DPV), electrochemical impedance spectroscopy (EIS) and scanning electron microscopy. The EIS and DPV responses of the biosensor were increased and decreased, respectively, by the DNA damage induced by BPA which led us to develop novel systems for detection of DNA damage. Our records confirmed that the biosensor was able to rapid and sensitive detection of DNA damage induced by BPA. Finally, according to the developed systems for detection of DNA damage, we have developed voltammetric and impedimetric methods for determination of BPA.