2019

Protein functionalised self assembled monolayer based biosensor for colon cancer detection.

Talanta Narayan T, Kumar S, Kumar S, Augustine S, Yadav BK, Malhotra BD
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We report results of the studies relating to the fabrication of a surface plasmon resonance (SPR) based label-free immunosensor for real-time monitoring of endothelin-1 (ET-1), a colon cancer biomarker. A gold disk modified with a self-assembled monolayer (SAM) of 11-mercaptoundecanoic acid (11-MUA) was functionalised via covalent immobilization of monoclonal anti-ET-1 antibodies using EDC-NHS (1-(3-(dimethylamine)-propyl)-3-ethylcarbodiimide hydrochloride, N-hydroxy succinimide) chemistry. This immunosensing platform (ethanolamine/anti-ET-1/11-MUA/Au) was characterized via atomic force microscopy (AFM), contact angle (CA) and Fourier transform infrared (FT-IR) spectroscopic techniques. The fabricated SPR electrode was further used to detect ET-1 in the broad concentration range 2-100 pg mL-1, with a detection limit of 0.30 pg mL-1 and remarkable sensitivity of 2.18 mo pg-1mL. The adsorption mechanism was studied using monophasic model and the values of association (ka) and dissociation (kd) constants for anti-ET-1 and ET-1 binding were calculated to be 4.4 ± 0.4 × 105 M-1 s-1 and 2.04 ± 0.0003 × 10-3 s-1, respectively. The results obtained via analysis of serum samples of colorectal cancer patients were found to be in good agreement with those obtained from enzyme-linked immunosorbent assay (ELISA) technique. Further, electrochemical studies were performed to prove the efficacy of the fabricated platform as a point of care device for the detection of ET-1.

关键词