2019

A Surface Acoustic Wave (SAW) biosensor method for functional quantification of E. colil-asparaginase.

Talanta Yao H, Fernández CS, Xu X, Wynendaele E, De Spiegeleer B
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Biosensors are rising technologies in the pharmaceutical field for medicine discovery, development and Quality Control (QC) stages. Surface acoustic wave (SAW) biosensor employs acoustic waves generated by oscillating a piezoelectric crystal quartz plate to meas. mass and viscosity, and allows to detect and quantify binding events between the analyte and an immobilized interacting ligand. We present here a SAW biosensor based approach for the functional quantification of Escherichia colil-asparaginase (E. colil-ASNase), using polyclonal antibody (pAb) as the interaction partner immobilized on the chip. Different immobilization strategies of pAb were initially evaluated, resulting in the BS3 activated amide coupling via protein G strategy as the final immobilization method. The method was validated by evaluating the selectivity, linearity, as well as accuracy (a recovery of 102.4%) and precision (RSD of 8.5%). The application of the validated method on different samples encompassing different lots of E. colil-ASNase, deamidated E. colil-ASNase and dry-heated E. colil-ASNase (80 °C, 10 min) indicated the suitability of the developed SAW method to quantify E. colil-ASNase. We suggest this SAW method can be adopted as a pharmaceutical QC method.

关键词