组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
Amperometric enzyme biosensors are some of the simplest and cheapest types of medical devices used in the rapid detection of biomarkers that have been developed in the past fifty years. When the concentrations of biomarkers are at micromoles per liter, such as for sarcosine, which was recently discovered as a biomarker for prostate cancer, the response signal of the interferences is huge, and the biosensor is hard to satisfy the requirements of practical applications. In this manuscript, we describe a strategy for synthesizing a surface electronegative organic-inorganic hybrid mesoporous material, which could reduce the interference signal much better than Nafion and Chitosan. We verify that the surface potential of the carrier nanomaterial plays an important role in excluding anionic interferences. We also prepare a sensitive (16.35 μA mM-1), low LOD (0.13 μM) and wide linear range (1-70 μM) amperometric sarcosine biosensor with excellent anti-interference properties. This mesoporous material provides a bio-composite platform for the development of simple amperometric biosensors for detecting micromoles per liter of analytes in serum or urine.