Detection of CD133-marked cancer stem cells by surface plasmon resonance: Its application in leukemia patients.

Biochimica et biophysica acta. General subjects Fathi F, Rahbarghazi R, Movassaghpour AA, Rashidi MR
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Here, we reported the development of a label-free and real-time surface plasmon resonance (SPR) based biosensor for cancer stem cells (CSCs) detection using cell surface biomarker; CD133. The fabricated biosensor was used for detection of this marker in some acute myeloid leukemia (AML) patients and the results were compared with those obtained from flow cytometry (FC) method. CD133 antibody was immobilized on the gold chip surface via EDC/NHS coupling method and binding of the candidate cells to the modified gold sensor surface was monitored after isolation of mononuclear cells from bone marrow of the patients. The method was validated in terms of various parameters such as CD133- antibody concentration and cell density. The CD133-marked cells were investigated in seven AML patients. All SPR results were compared with those obtained from FC method. A very good correlation (R2 = 0.96) was obtained between SPR and FC responses related to CD133-marked cells densities. In conclusion, in this study, a label-free and real-time SPR cytometry method was developed to detect CD133 and it was successfully applied to follow this cancer stem cell biomarker in AML patients.