2019

Detection of

Micromachines Koo B, Jin CE, Bae M, Jang YO, Kim JY, Kim SH, Shin Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Blood plasma from patients is a powerful resource for diagnosing infectious disease due to it having many genetic materials as well as being relatively easy to obtain. Thus, various biosensors have been investigated for diagnosing diseases in blood plasma. However, there are no optimized and validated sensors for clinical use due to the low sensitivity, complexity, and difficulties of removing the inhibitors from plasma samples. In this study, we described a silicon microring resonator sensor used to detect Coxiella burnetii from the blood plasma of Q-fever patients in a label-free, real-time manner. Q-fever is an infectious disease caused by Coxiella burnetii via direct contact or inhalation aerosols. We validated this biosensor in the blood plasma of 35 clinical samples (including 16 Q fever samples infected with Coxiella burnetii and 19 samples infected with other febrile diseases. The biosensors are capable of rapid (10 min), highly sensitive (87.5%), and specific (89.5%) detection in plasma samples compared to the use of the conventional method.

关键词