Self-build packed-bed bioreactor for rapid and effective BOD estimation.

Environmental science and pollution research international Pham TTP, Nguyen PHD, Nguyen TTV, Duong HTL
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This work demonstrated a simple, low-cost, rapid, and effective biochemical oxygen demand (BOD) estimation system based on a packed-bed bioreactor that can be easily self-built on-site at a particular wastewater treatment plant for continuous monitoring of the influent and effluent. The use of natural microbial consortium that were collected from the target wastewater and immobilized on a cheap porous carrier simply by adhesion resulted in an acceptable accuracy of over 95%. The newly developed semi-continuous operating mode with peak-type signals was shown to be able to continuously estimate BOD at a high flow rate to overcome the flow dependence of the oxygen electrode, limit clogging issues, enhance the response time, and lower the limit of detection. The resulting packed-bed bioreactors could work continuously for 22 h with a coefficient of variance (CoV) of only 1.8% or for 13 h a day for several days with a maximum CoV of 1.4% and their response was observed to be stable over 80 consecutive measurements. They exhibited stable responses at a wide pH range of 6.5-8.5, which is also the recommended range for aerobic wastewater treatment, emphasizing the greater ease of use of natural microorganisms for BOD estimation.