In this paper, a novel label-free electrochemical impedance aptasensor for highly sensitive detection of IFN-γ based on target-induced exonuclease inhibition was constructed. For this purpose, we designed a DNA hairpin modified on the gold electrode whose loop was the aptamer of the IFN-γ, and the stem was 5'-thiol-modified. In the absence of IFN-γ, Exonuclease III (Exo III) and Exonuclease I (Exo I) digested the double-stranded and single-stranded strands of the hairpin DNA, respectively, causing smaller impedance value on the surface of the electrode. In the presence of IFN-γ, the function of Exo III was greatly inhibited by the binding of the aptamer with the target, and it stopped after cutting three bases of the hairpin DNA. Forming a major target-bound aptamer digestion product, it could not be digested by Exo I, so there was larger impedance on the electrode surface. The calibration curve for IFN-γ was linear in the range of 1 pM-50 nM with the detection limit (LOD) of 0.7 pM. The proposed aptasensor proved good selectivity and reproducibility, and low cost. In addition, the biosensor was able to detect IFN-γ in serum samples successfully, which is expected to provide an efficient method for TB diagnosis at early stages.