2019

Highly Sensitive Endotoxin Assay Combining Peptide/Graphene Oxide and DNA-Modified Gold Nanoparticles.

ACS omega Yu N, Zhang X, Gao Y, You H, Zhang J, Miao P
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Endotoxin is a highly toxic stimulator originated from the outer membrane of Gram-negative bacteria, which should be monitored sensitively and selectively for human health concerns. Traditional detection methods mainly rely on limulus amoebocyte lysate assay. However, it suffers drawbacks like the narrow detection range, and the results may be environment-dependent. In this work, we have developed a sensitive electrochemical biosensor for endotoxin assay. Peptide is first designed as specific recognition element toward endotoxin. Graphene oxide and DNA-modified gold nanoparticles are then used to enhance the electrochemical signal. The analytical performances are excellent with the limit of detection as low as 0.001 EU mL-1. This method has also been successfully applied in endotoxin assay in complex biological samples, which may have great potential use.

关键词