2019

Microvesicle detection by a reduced graphene oxide field-effect transistor biosensor based on a membrane biotinylation strategy.

The Analyst Wu D, Zhang H, Jin D, Yu Y, Pang DW, Xiao MM, Zhang ZL, Zhang ZY, Zhang GJ
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Unlike other extracellular vesicle (EV) subtypes such as exosomes, the lack of well-defined universal markers on the surface of microvesicles (MVs) has led to difficulty in the detection of the entire MV population. To design a universal MV detection method, we reported highly sensitive electrical detection of MVs using a reduced graphene oxide (RGO)-based field-effect transistor (FET) biosensor by the introduction of a membrane biotinylation strategy in this work. Biotinylated MVs (B-MVs) were obtained by supplying the culture medium with 1,2-distearoyl-sn-glycero-3-phosphoethanolamine-N-[biotinyl(polyethylene glycol)-2000] (DSPE-PEG-biotin) while cultivating the cells. Excellent biotinylation efficiency of MVs (92.6%) was then realized. A streptavidin (SA) probe was subsequently modified onto the channel surface of the as-fabricated RGO-based FET device, which was capable of specifically recognizing B-MVs due to the high affinity between SA and biotin in a 1 : 4 recognition format. The results showed that the RGO-based FET biosensor could detect B-MVs in a wide range from 105 particles per mL to 109 particles per mL with a low detection limit down to 20 particles per μL, which was the lowest value compared with other previously reported results. This platform also allowed distinguishing B-MVs from other unbiotinylated EV types such as MVs and exosomes, exhibiting excellent specificity. Moreover, this FET biosensor demonstrated the capability of detecting B-MVs derived from different cell lines including cancer cells and normal cells, indicating its versatility and potential applications in the biomedical field.

关键词