组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
A critical barrier for the successful development of fiber sensors for bio-chemical processes is their limitedly improved sensitivity, restricted by the sensor structural design. To solve this, in this paper, a novel concept was proposed using functionalised modified magnetic microspheres (MMSs) to "amplify" the effect of target bio-chemical analytes to significantly improve the fiber sensor's sensitivity, which has been demonstrated using human chorionic gonadotropin (hCG) as an example. Two types of antibody hCG, (β and α, both can specifically bind with hCG), were adhered on the surface of fibre sensor and MMSs respectively. Both hCG and MMSs will be specifically captured by the fibre sensor, where MMSs act as an "amplifier" to improve the sensor sensitivity. Experimentally immunomagnetic detection limit of 0.0001 mIU/mL has been achieved, which is the highest reported so far. This newly developed methodology opens a new direction for sensitivity improvement and could be further explored to applications require ultrahigh sensitivity detections such as earlier medical diagnostics.