2019

Microcantilever Array Biosensor for Simultaneous Detection of Carcinoembryonic Antigens and α-Fetoprotein Based on Real-Time Monitoring of the Profile of Cantilever.

ACS sensors Li C, Ma X, Guan Y, Tang J, Zhang B
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A microcantilever array biosensor based on a sandwich structure has been developed for simultaneously measuring two biomarkers carcinoembryonic antigen (CEA) and α-fetoprotein (AFP) via an optical readout technique-real-time monitoring of the profile of cantilever. First, the aptamers of CEA and AFP were self-assembled on their respective cantilevers. After the adsorption of the mixture of CEA and AFP, further specific interaction was performed via the addition of the antibodies specific to each target. The compressive stress on the cantilever was generated by the aptamer-antigen-antibody sandwich structure formed on the gold surface, resulting in cantilever bending. The profile of cantilever could be monitored in real time. The relationship between the deflection value at the 90% position of the cantilever and the target concentration served as a calibration curve, and the detection sensitivity was 1.3 ng/mL for CEA and 0.6 ng/mL for AFP, respectively. This work demonstrated the ability of simultaneously measuring two biomarkers via a microcantilever array biosensor, giving great potential for further application in detecting several targets simultaneously for early clinical diagnosis.