2019

Development Toward a Triple-Marker Biosensor for Diagnosing Cardiovascular Disease.

Critical reviews in biomedical engineering Deng A, Matloff D, Lin CE, Probst D, Broniak T, Alsuwailem M, La Belle JT
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Cardiovascular disease (CVD) is the leading cause of death in the United States and is responsible for 30% of all deaths globally. The diagnosis and management of CVD requires monitoring of multiple biomarkers, which comprehensively represents the state of the disease. However, many assays for cardiac biomarkers today are complicated and laborious to perform. Rapid and sensitive biosensors capable of giving accurate measurements of vital cardiac biomarkers without complex procedures are thus in high demand. In the work presented below, rapid, label-free biosensor prototypes for three Food and Drug Administration-approved biomarkers are reported: B-type natriuretic peptide (BNP), cardiac troponin I (cTnI), and C-reactive protein (CRP). The sensors were prepared by immobilizing each biomarker's antibody onto gold working electrodes with platinum counter and silver/silver chloride reference electrodes. The sensors were tested using electrochemical impedance spectroscopy (EIS), a femto-molar sensitive technique capable of label-free, multi-marker detection if a biomarker's optimal frequency (OF) can be identified. The OFs of BNP, cTnI, and CRP were found to be 1.74, 37.56, and 253.9 Hz, respectively. The performance of the BNP biosensor was also evaluated in blood and achieved clinically relevant detection limits of 100 pg/mL.