Toward a Label-Free Electrochemical Impedance Immunosensor Design for Quantifying Cortisol in Tears.

Critical reviews in biomedical engineering Cardinell BA, Spano ML, La Belle JT
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Cortisol is a viable biomarker for monitoring physiological, occupational, and emotional stress and is normally present in tear fluid at approximately 40 nM, or higher as a result of stress. We present characterization and quantification of cortisol via several electrochemical methods versus the standard enzyme-linked immunosorbent assay, commonly known as ELISA. We also present a prototyped design of a disposable test strip and handheld sensor based on label-free electrochemical impedance spectroscopy to quantify cortisol levels in tear fluid within approximately 90 seconds. Electrochemical characterization of the cortisol molecule was conducted using cyclic voltammetry, amperometric i-t, and square wave voltammetry. Lower limits of detection for these techniques were not sufficient to quantify cortisol and phycological tear ranges: 0.1 M, 0.23 M, and 193 M for cyclic voltammetry, amperometric i-t, and square wave voltammetry, respectively. However, electrochemical impedance spectroscopy (EIS) was to be the best mode of cortisol quantification and comparison to ELISA technique (detection range of ~ 138 pM - 552 nM). The initial EIS biosensor obtained a lower limit of detection of 59.76 nM with an approximate 10% relative standard deviation. The cortisol assay and tear collection prototype presented here offer a highly reproducible and ultra-low level of detection with a label-free and rapid response.