Clinically comparable impedimetric immunosensor for serum alkaline phosphatase detection based on electrochemically engineered Au-nano-Dendroids and graphene oxide nanocomposite.

Biosensors & bioelectronics Mahato K, Purohit B, Kumar A, Chandra P
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work, we demonstrate label-free electrochemical impedance spectroscopy (EIS) based alkaline phosphatase (ALP) detection using gold nanoparticles (AuNPs), electrochemically engineered Au-nano-Dendroids, and graphene oxide (GO) nanocomposite. These nanomaterials were sequentially deposited on to the screen-printed carbon electrode (SPCE) and antibodies against ALP (anti-ALP) were immobilized using carbodiimide bioconjugation process. The sensor probe has been characterized extensively using TEM, EDX, SAED, XRD, FE-SEM, FTIR, DIC, and electrochemical techniques. The analytical performance of fabricated biosensor has been evaluated using EIS, where linear dynamic range and limit of detection were obtained to be 100-1000 U/L and 9.10 (±0.12) U/L, respectively. The developed biosensor showed high selectivity towards ALP with negligible interference (ksel « 1; n = 3) due to coexisting molecules. The sensor probe has successfully recovered ALP between 108.84% and 172.50% (n = 3) in human serum samples. The sensor has been used to estimate ALP in clinical serum samples, where the level was found to be 83.15 U/L and was comparable with standard technique used in the hospitals. The shelf life, stability, and reproducibility have also been evaluated.