组成图示
示意图生成中
传感器类型
—
检测对象
—
检测原理
—
检测灵敏度
—
效应效果
—
传感器的构成
—
中文摘要
—
英文摘要
In this paper, we present a highly sensitive and selective detection of serum carcinoembryonic antigen (CEA) based on silicon nanowire (SiNW) array device. With the help of traditional microfabrication technology, low-cost and highly controllable SiNW array devices were fabricated. After a series of surface modification processes, SiNW array biosensors show rapid and reliable response to CEA; the detection limit of serum CEA was 10 fg/mL, the current signal is linear with the logarithm of serum CEA concentration in the range of 10 fg/mL to 100 pg/mL. In this work, SiNW array biosensors can obtain strong signal and high signal-to-noise ratio; these advantages can reduce the production cost of the SiNW-based system and promote the application of SiNWs in the field of tumor marker detection.