2019

Characterization of an Aptamer Directed against 25-Hydroxyvitamin D for the Development of a Competitive Aptamer-Based Assay.

Biosensors Prante M, Schüling T, Roth B, Bremer K, Walter J
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Detection of the small molecule 25-hydroxyvitamin D (25(OH)D) as the most relevant marker for vitamin D supply suffers from a high variability of results using the current detection methods, such as high-performance liquid chromatography (HPLC) and immunoassays. A new detection approach using a highly specific aptamer directed against 25(OH)D was established in this study based on the target-induced dissociation (TID) sensing approach. In this work, the aptamer was investigated regarding its structural properties as well as its binding affinity by using microscale thermophoresis (MST). Moreover, complementary oligonucleotides were designed based on the aptamer structure and were evaluated in MST experiments. Binding experiments of immobilized aptamers were conducted in microarray experiments. It could be shown that the aptamer exhibited the usual B-DNA structure and did not form any G-quadruplexes. The design of complementary oligonucleotides for the TID assay identified a putative 25(OH)D binding site within the aptamer. The limit of detection of the established competitive assay was determined to be 5.4 nM, which sets the stage for the development of a biosensor system.

关键词