Quantum dots as nanolabels for breast cancer biomarker HER2-ECD analysis in human serum.

Talanta Freitas M, Neves MMPS, Nouws HPA, Delerue-Matos C
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Early detection of cancer increases the possibility for an adequate and successful treatment of the disease. Therefore, in this work, a disposable electrochemical immunosensor for the front-line detection of the ExtraCellular Domain of the Human Epidermal growth factor Receptor 2 (HER2-ECD), a breast cancer biomarker, in a simple and efficient manner is presented. Bare screen-printed carbon electrodes were selected as the transducer onto which a sandwich immunoassay was developed. The affinity process was detected through the use of an electroactive label, core/shell CdSe@ZnS Quantum Dots, by differential pulse anodic stripping voltammetry in a total time assay of 2 h, with an actual hands-on time of less than 30 min. The proposed immunosensor responded linearly to HER2-ECD concentration within a wide range (10-150 ng/mL), showing acceptable precision and a limit of detection (2.1 ng/mL, corresponding to a detected amount (sample volume = 40 μL) of 1.18 fmol) which is about 7 times lower than the established cut-off value (15 ng/mL). The usefulness of the developed methodology was tested through the analysis of spiked human serum samples. The reliability of the presented biosensor for the selective screening of HER2-ECD was confirmed by analysing another breast cancer biomarker (CA15-3) and several human serum proteins.