A novel 3D paper-based microfluidic electrochemical glucose biosensor based on rGO-TEPA/PB sensitive film.

Analytica chimica acta Cao L, Han GC, Xiao H, Chen Z, Fang C
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A novel 3D paper-based microfluidic screen-printed electrode (SPE) composed of two layers was constructed by photolithography and screen-printing technology. Aldehyde functionalized hydrophilic zone of the counter and reference electrodes layer was prepared for glucose oxidase immobilization. Highly conductive prussian blue deposited reduced graphene oxide-tetraethylene pentamine (rGO-TEPA/PB) modified paper working electrode layer can be used as an electrochemical sensitive membrane for quantitative detection of hydrogen peroxide (H2O2), which was the enzyme-catalyzed reaction product. Therefore, this 3D paper-based microfluidic electrochemical biosensor can be used for quantitative detection of glucose. Under optimum conditions, the proposed biosensor can be used for quantitative determination of glucose over a wide linear range of 0.1 mM-25 mM with detection limit of 25 μM. Finally, the 3D paper-based microfluidic electrochemical biosensor was applied to determine glucose in human sweat and blood, and the obtained results were in good consistency with values measured by Roche's blood glucose meter. In addition, the proposed 3D paper-based electrochemical device showed good repeatability, stability, and anti-interference, which would be of great potential to monitor glucose in complex biological fluids.