Comparing a Distributed Parameter Model-Based System Identification Technique with More Conventional Methods for Inverse Problems.

Journal of inverse and ill-posed problems Li J, Luczak SE, Rosen IG
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Three methods for the estimation of blood or breath alcohol concentration (BAC/BrAC) from biosensor measured transdermal alcohol concentration (TAC) are evaluated and compared. Specifically, we consider a system identification/quasi-blind deconvolution scheme based on a distributed parameter model with unbounded input and output for ethanol transport in the skin and compare it to two more conventional system identification and filtering/deconvolution techniques for ill-posed inverse problems, one based on frequency domain methods, and the other on a time series approach using an ARMA input/output model. Our basis for comparison are five statistical measures of interest to alcohol researchers and clinicians: peak BAC/BrAC, time of peak BAC/BrAC, the ascending and descending slopes of the BAC/BrAC curve, and the area underneath the BAC/BrAC curve.