2020

Copper ion modified graphitic C

Spectrochimica acta. Part A, Molecular and biomolecular spectroscopy Cao JT, Zhang WS, Fu XL, Wang H, Ma SH, Liu YM
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A high-efficient chemiluminescence (CL) platform for highly selective and sensitive H2S detection was constructed on the basis of the quenching effect of S2- on the copper ion modified graphitic carbon nitride nanosheets (Cu2+-g-C3N4 NSs) enhanced luminol-H2O2 system. Cu2+-g-C3N4 NSs with horseradish peroxidase-like catalytic activity were prepared and provide a great improvement for luminol-H2O2 system. The presence of S2- induced the formation of CuS precipitate on g-C3N4 NSs surface. The precipitate can block the catalytic Cu2+ sites on the g-C3N4 NSs surface, resulting in a great CL decrease of CL system. Based on such a mechanism, a simple, highly selective and sensitive CL biosensor for H2S detection was designed. Under the optimized conditions, luminol-H2O2-Cu2+-g-C3N4 NSs system gave a decrease of CL intensity with the Na2S concentration increasing. The CL biosensor is in a linear range of 10.0 pM-50.0 nM and the detection limit for detecting Na2S is as low as 2.0 pM. Moreover, the method here has enjoyed a successful application for determining H2S in human plasma samples and the recovery is between 95.7% and 110.0%.

关键词