2019

Compressive Sensing of Cuff-less Biosensor for Energy-Efficient Blood Pressure Monitoring.

Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference Rachim VP, Chung WY
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Compressive Sensing (CS) is an emerging technique in Internet of Medical Things (IoMT) application especially for smart wearable devices to prolong the sensor lifetime, and enable a continuous healthcare monitoring system. This paper describes the performance of CS on our wrist-based cuff-less biosensor for estimating blood pressure (BP) continuously. The proposed biosensor offers a novel BP estimation method by only using the biosignal from subject wrist. A CS technique is implemented to encrypt and reduce the data transmission load of the dual biosignal, which include impedance plethysmography (IPG) and photo plethysmography (PPG). Therefore, multiple compression ratio (CR) were tested to the original signal. We further compare the CS-based extracted features called pulse transit times (PTTs). Based on our experiments, CS-based PTT value that calculated from the IPG peak point to the PPG max2 point (F2), achieved the best correlation coefficient (R) of -0.85 and -0.43 to the systolic BP and diastolic BP, respectively. These results suggest that the implementation of CS on our proposed wrist biosensor is suitable for non-intrusive, yet long-term continuous BP monitoring.

关键词