2020

An antifouling peptide-based biosensor for determination of Streptococcus pneumonia markers in human serum.

Biosensors & bioelectronics Chang PH, Weng CC, Li BR, Li YK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We report a peptide-based sensor that involves a multivalent interaction with L-ascorbate 6-phosphate lactonase (UlaG), a protein marker of Streptococcus pneumonia. By integrating the antifouling feature of the sensor, we significantly improved the signal-to-noise ratio of UlaG detection. The antifouling surface was fabricated via electrodeposition using an equivalent mixture of 4-amino-N,N,N-trimethylanilinium and 4-aminobenzenesulfonate. This antifouling layer not only effectively reduces the non-specific adsorption on the biosensor but also decreases the charge transfer resistance (Rct) of the screen-printed carbon electrode. The aniline-modified S7 peptide, an UlaG-binding peptide, was pre-synthesized and further electrochemically modified to bind onto the antifouling layer. Bio-electrochemical analysis confirms that the antifouling S7-peptide sensor binds strongly to the UlaG with a dissociation constant (Kd) = 0.5 nM. This strong interaction can be attributed to a multivalent interaction between the biosensor and the heximeric form of UlaG. To demonstrate the potential for clinical application, further detection of Streptococcus pneumonia from 50 to 5×104 CFU/mL were successfully performed in 25% human serum.