2020

Modification of reduced graphene/Au-aptamer to develop an electrochemical based aptasensor for measurement of glycated albumin.

Talanta Farzadfard A, Shayeh JS, Habibi-Rezaei M, Omidi M
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Herein, an electrochemical label-free biosensor designed for the detection of glycated albumin (GA) using reduced graphene oxide/Au nanoparticles (RGO/AuNPs) modified by anti-GA aptamer. For fast and simple modification of the electrode, the aptamer chain was thiolated. Transition electron microscopy (TEM), Fourier transform infrared spectroscopy (FTIR), X-ray diffraction (XRD) techniques were used to the characterization of synthesized materials. Structural analysis of nanomaterials shows that graphene sheets were synthesized very fine by average thickness of 2.5 nm and Au nanoparticles distributed on the surface of graphene sheets uniformly. Cyclic voltammetry (CV) square wave voltammetry (SWV) and impedance spectroscopy (EIS) were used to electrochemical study of the decorated electrode. Electrochemical studies described the potential of fabricated rGO/AuNPs-aptamer electrode to selectively determine GA properly in buffer solution at the range of 2-10 μg mL-1 by the detection limit of 0.07 μg. mL-1 for GA.

关键词