2020

Highly Sensitive Closed Loop Enclosed Split Ring Biosensor With High Field Confinement for Aqueous and Blood-Glucose Measurements.

Scientific reports Kandwal A, Igbe T, Li J, Liu Y, Li S, Liu LWY, Nie Z
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

This paper presents a highly sensitive closed loop enclosed split ring biosensor operating in microwave frequencies for measuring blood glucose levels in the human body. The proposed microwave glucose biosensor, working on the principle of high field confinement and concentrated energy, has been tested using both in-vitro and in-vivo methods. This principle allows the sensor to concentrate energy at the surface which results in improved accuracy of measurements. For in-vitro measurements, the biosensor has been tested using de-ionized water glucose solutions of different concentrations. The miniaturized micrometer scale biosensor is fabricated over a thin Si-substrate using photolithographic technique. The biosensor has been designed in a way to operate at desired microwave frequencies. Highly confined fields and concentrated energy inside the closed loop line containing the split ring resonators are responsible for the sensitivity enhancement. This new biosensor has obtained a high sensitivity of 82 MHz/mgmL-1 within the clinical diabetic range during in-vivo testing over the human body. In addition, the subjects (undergoing experiments) steady state has been continuously monitored throughout the experiment which helps in improving the accuracy of the results. The proposed biosensor has further obtained a low detection limit of <0.05 wt.% and can be useful for continuous non-invasive blood glucose monitoring.

关键词