Highly sensitive electrochemiluminescence biosensor for VEGF

Analytical and bioanalytical chemistry Cheng JL, Liu XP, Chen JS, Mao CJ, Jin BK
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this work, an electrochemiluminescence (ECL) biosensor was fabricated for the selective detection of vascular endothelial growth factor (VEGF165). g-C3N4/PDDA/CdSe nanocomposites were used as the ECL substrate. Then, DNA labeled at the 5' end with amino groups (DNA1) was immobilized on the surface of g-C3N4/PDDA/CdSe nanocomposite-modified glassy carbon electrode (GCE) by amido linkage. AuNP-labeled target DNA (Au-DNA2) could hybridize with DNA1 to form a double strand. The ECL of the g-C3N4/PDDA/CdSe nanocomposite was efficiently quenched due to the resonance energy transfer between CdSe QDs and Au NPs. After VEGF165 was recognized and bound by Au-DNA2, the double helix was disrupted, and the energy transfer was broken. In this case, Au-DNA2 was released from the electrode surface, and the ECL intensity recovered to a higher level. Under optimal conditions, this ECL biosensor possesses excellent selectivity, accuracy, and stability for VEGF165 detection in a linear range of 2 pg mL-1 to 2 ng mL-1 with a detection limit of 0.68 pg mL-1. In addition, this assay has been successfully applied to the determination of VEGF165 in serum samples. Graphical abstract Schematic representation of the electrochemiluminescence sensor based on a g-C3N4/PDDA/CdSe nanocomposite, which can be determined in the concentration of vascular endothelial growth factor in serum.