2020

Electrochemical immunoassay for the detection of stress biomarkers.

Heliyon Abdulsattar JO, Greenway GM, Wadhawan JD
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A rapid electrochemical immunoassay method was developed to detect and measure stress biomarkers (cortisol and cortisone) in two biological samples (Zebrafish whole-body and artificial saliva). This methodology utilizes an immunoassay approach taking advantage of the lock and key mechanism that is related to the antibody-antigen interaction depending on the reliable immobilization of the antibody labelled with ferrocene tags (Ab-Fc) on a modified tin-doped indium oxide (ITO) electrode using electrochemical instrumentation to build a POC platform. The limit of detection (LOD) obtained for this biosensor was 1.03 pg ml-1 for cortisol and 0.68 pg ml-1 for cortisone, respectively. The correlation coefficient was 0.9852 and 0.9841 for cortisol and cortisone, respectively with a linear concentration from (0-50 ng ml-1) which covers the standard levels of stress hormones in both selected biological samples. The incubation time was investigated and 30 min was found to be the optimum incubation time. This time would be acceptable for the POC system as total process time can be determined within 35 min.

关键词