Development of a highly sensitive, field operable biosensor for serological studies of Ebola virus in central Africa.

Sensors and actuators. B, Chemical Petrosova A, Konry T, Cosnier S, Trakht I, Lutwama J, Rwaguma E, Chepurnov A, Mühlberger E, Lobel L, Marks RS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We describe herein a newly developed optical immunosensor for detection of antibodies directed against antigens of the Ebola virus strains Zaire and Sudan. We employed a photo immobilization methodology based on a photoactivatable electrogenerated poly(pyrrole-benzophenone) film deposited upon an indium tin oxide (ITO) modified conductive surface fiber-optic. It was then linked to a biological receptor, Ebola virus antigen in this case, on the fiber tip through a light driven reaction. The photochemically modified optical fibers were tested as an immunosensor for detection of antibodies against Ebola virus, in animal and human sera, by use of a coupled chemiluminescent reaction. The immunosensor was tested for sensitivity, specificity, and compared to standard chemiluminescent ELISA under the same conditions. The analyte, anti-Ebola IgG, was detected at a low titer of 1:960,000 and 1:1,000,000 for subtypes Zaire and Sudan, respectively. While the same serum tested by ELISA was one order (24 times) less sensitive.

关键词