Accuracy improvement of electrochemical whole blood ketone sensor based on HCT compensation algorithm.

Electrophoresis Zhao Y, Gu H, Li H, Huang L, Cen X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In this manuscript, we demonstrated an electrochemical test strip with HCT (hematocrit) compensation algorithm to improve the accuracy of blood ketone sensor. In the conventional electrochemical sensor, the electrochemical current was directly resolved into the concentration value of the determinant without HCT compensation. For lower or higher HCT blood sample, the measured result was inaccurate. In the proposed design, the blood impedance can be measured to estimate the HCT, which was utilized to compensate the electrochemical current to resolve the more accurate concentration of determinant. The practical blood sample tests demonstrated the proposed design can provide more believable and reliable measured result in clinical point-of-care setting.