The progesterone (P4) level in body fluids can act as an indicator for early pregnancy diagnosis and offers insight into mammalian somatic function. In this work, we designed an antibody-aptamer based sandwich assay as a cathodic photoelectrochemical (PEC) biosensor for P4 detection. The composites of carbon dots and graphene oxide (CDs-GO) with favorable cathodic photocurrent response were used as photoactive materials on which the antibody (Ab) of P4 was immobilized. Meanwhile, high affinity truncated P4 aptamer was immobilized on Au-CuO-Cu2O to act as a bioconjugate. When P4 was present, the aptamer-Au-CuO-Cu2O bioconjugate could amplify the cathodic photocurrent of CDs-GO modified electrode through Ab-P4-aptamer interactions. Under optimum conditions, the cathodic photocurrent of the constructed PEC biosensor was found to increase linearly with P4 in a wide concentration range from 0.5 nM to 180 nM, with a low detection limit (3S/N) of 0.17 nΜ. The proposed cathodic PEC sensing platform demonstrated high selectivity, satisfying reproducibility, good stability. The sensor was successfully applied in the determination of P4 in human serum samples.