2020

10.3390/bios10060064

Biosensors Hashemi Tameh M, Primiceri E, Chiriacò MS, Poltronieri P, Bahar M, Maruccio G
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Pectobacterium atrosepticum (Pba) is a quarantine and threatening phytopathogen known as the causal agent of blackleg and soft rot disease of potatoes in many areas. Its early detection is then important to have healthy potato tubers and reduce economic losses. Today, conventional methods such as enzyme-linked immunosorbent-assay (ELISA) and polymerase chain reaction (PCR) are typically used for Pba detection, but they are expensive and time-consuming. Here we report on the optimization of an alternative approach based on an electrochemical impedance immunosensor combining a microfluidic module and a microelectrodes array, and having advantages in terms of low cost, ease of use and portability. For validation and for assessing its performance, the lab-on-chip platform has been compared with two standard methods (ELISA and PCR).

关键词