2020

High-throughput imaging surface plasmon resonance biosensing based on ultrafast two-point spectral-dip tracking scheme.

Optics express Zeng Y, Wang X, Zhou J, Miyan R, Qu J, Ho HP, Zhou K, Gao BZ, Chen J, Shao Y
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Wavelength interrogation surface plasmon resonance imaging (λSPRi) has potential in detecting 2-dimensional (2D) sensor array sites, but the resonance wavelength imaging rate limits the application of detecting biomolecular binding process in real time. In this paper, we have successfully demonstrated an ultrafast λSPRi biosensor system. The key feature is a two-point tracking algorithm that drives the liquid crystal tunable filter (LCTF) to achieve fast-tracking of the resonance wavelength movement caused by the binding of target molecules with the probe molecules on the sensing surface. The resonance wavelength measurement time is within 0.25s. To date, this is the fastest speed ever reported in λSPRi. Experiment results show that the sensitivity and dynamic are 2.4 × 10-6 RIU and 4.6 × 10-2 RIU, respectively. In addition, we have also demonstrated that the system has the capability of performing fast high-throughput detection of biomolecular interactions, which confirms that this fast real-time detecting approach is most suitable for high-throughput and label-free biosensing applications.

关键词