2020

Plasmonic/magnetic molybdenum trioxide and graphitic carbon nitride quantum dots-based fluoroimmunosensing system for influenza virus.

Sensors and actuators. B, Chemical Achadu OJ, Takemura K, Khoris IM, Park EY
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A novel magnetic/plasmonic-assisted fluoro-immunoassay system is developed for the detection of influenza virus using magnetic-derivatized plasmonic molybdenum trioxide quantum dots (MP-MoO3 QDs) as the plasmonic/magnetic agent and fluorescent graphitic carbon nitride quantum dots (gCNQDs) as the monitoring probe. Specific antibody against influenza A virus was conjugated onto the surface of MP-MoO3 QDs and gCNQDs, respectively. In the presence of influenza A virus (as the test virus), a core-satellite immunocomplex is formed between the antibody-conjugated nanomaterials (Ab-MP-MoO3 QDs and Ab-gCNQDs) and their interaction resulted in the modulation and gradual enhancement of the fluorescence intensity of the detection probe with the influenza virus concentration-dependent increase. In addition, PL change without influenza A virus was not observed. Limits of detection of 0.25 and 0.9 pg/mL were achieved for Influenza virus A/New Caledonia (20/99/IVR/116) (H1N1) detection in deionized water and human serum, respectively. Clinically isolated influenza virus A/Yokohama (110/2009) (H3N2) was detected in the range of 45 - 25,000 PFU/mL, with a limit of detection ca 45 PFU/mL (as opposed to a minimum of 5000 PFU/mL for a commercial test kit). This developed biosensor provides a robust, sensitive as well as a selective platform for influenza virus detection.