2020

Three-dimensional biosensor surface based on novel thorns-like polyelectrolytes.

Biosensors & bioelectronics Pan W, Han Z, Chang Y, Duan X
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

We have developed a new three-dimensional (3D) surface for use in biosensors that is based on modified novel thorns-like polyelectrolytes (3D-PETx), which comprises of poly-L-lysine (PLL) appended with multitude oligo (ethylene glycol) (OEG) and biotin moieties. It tethered to the sensor surface by PLL, while the OEG-biotin chains are forced to stretch away from the surface for target detections. Due to its 3D structure, the number of the OEG-biotin per surface unit is markedly increased compared to conventional 2D polyelectrolytes (2D-PET) coating. Their antifouling property and sensing performance for human IgG and PSA were compared with the 2D-PET by BioLayer Interferometry (Blitz), Surface Plasmon Resonance (SPR), microfluidic devices and Enzyme-Linked ImmunoSorbent Assay (ELISA). Experimental results show that 3D-PETx presents higher sensitivity for biomarker detection both in buffer and in serum and provides an almost non-fouling surface even in undiluted serum. In addition, a sensitive PSA detection was achieved in undiluted serum with a LOD down to 0.6 ng/mL. The successful immunosensing in undiluted serum demonstrate the potential of the 3D-PETx coating for real clinical applications.

关键词