2020

Construction of a novel electrochemical biosensor based on a mesoporous silica/oriented graphene oxide planar electrode for detecting hydrogen peroxide.

Analytical methods : advancing methods and applications Lu KC, Wang JK, Lin DH, Chen X, Yin SY, Chen GS
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

A constant magnetic field (CMF) was used to arrange the orientation of graphene oxide (GO) which was modified on a self-made screen-printed electrode. We evaluated the efficiency of this method for potential analytical application towards the sensing of hydrogen peroxide (H2O2). Mesoporous silica (MS)-encapsulated horseradish peroxidase (HRP) was immobilized on the electrode with vertically arranged GO to construct an H2O2 sensor (denoted as CMF/GO/HRP@MS). The linear range of the response of the CMF/GO/HRP@MS sensor to H2O2 was 0.1-235 μM, and the detection limit was as low as 0.01 μM. The results demonstrated that the vertical arrangement of GO resulting from the CMF on the electrode surface could increase the electron transfer rate. The excellent selectivity and anti-interference ability of this sensor to H2O2 in physiological samples may be attributed to the synergistic effect of mesoporous silica, GO and constant magnetic field.