Retrospective clinical evaluation of 4 lateral flow assays for the detection of SARS-CoV-2 IgG.

Diagnostic microbiology and infectious disease McAulay K, Bryan A, Greninger AL, Grill F, Lake D, Kaleta EJ, Grys TE
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

In a Clinical Laboratory Improvement Amendments laboratory setting, we evaluated severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) IgG detection with 4 lateral flow immunoassays [LFIAs; 2 iterations from BTNX Inc. (n = 457) and 1 each from ACON Laboratories (n = 200) and SD BIOSENSOR (n = 155)]. In a cohort of primarily hospitalized, reverse-transcription polymerase chain reaction-confirmed coronavirus disease 2019 cases, sensitivity at ≥14 days from symptom onset was: BTNX kit 1, 95%; BTNX kit 2, 91%; ACON, 95%; and SD, 92%. All assays showed good concordance with the Abbott SARS-CoV-2 IgG assay at ≥14 days from symptom onset: BTNX kit 1, 99%; BTNX kit 2, 94%; ACON, 99%; and SD, 100%. Specificity, measured using specimens collected prior to SARS-CoV-2 circulation in the United States and "cross-reactivity challenge" specimens, was 98% for BTNX kit 1 and ACON and 100% for BTNX kit 2 and SD. These results suggest that LFIAs may provide adequate results for rapid detection of SARS-CoV-2.