2020

Design of a structure-based fluorescent biosensor from bioengineered arginine deiminase for rapid determination of L-arginine.

International journal of biological macromolecules Tam SY, Chung SF, Chen YW, So YH, So PK, Cheong WL, Wong KY, Leung YC
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Rationally designed mutations on recombinant arginine deiminase (ADI) could act as a 'turn-off' L-arginine (L-Arg) fluorescent biosensor and provide an alternative method for rapid determination of L-Arg. Double mutations were introduced on the Cys251➔Ser251 and Thr265➔Cys265 of recombinant ADI, rendering a single cysteine present on the protein surface for the site-specific attachment of a fluorophore, fluorescein-5-maleimide. The double mutations on ADI (265C) and its fluorescein-labelled form (265Cf) conserved the catalytic efficiency of wild-type ADI. Upon binding to L-Arg, 265Cf induced structural conformational changes and rendered the fluorescein moiety to move closer to Trp264, resulting in fluorescence quenching. The duration of fluorescence quenching was dependant on the L-Arg concentration. A linear relationship between the time at the maximum rate of fluorescence change and L-Arg concentrations, which ranged from 2.5 to 100 μM, was found with R2 = 0.9988. The measurement time was within 0.15-4 min. Determination of L-Arg concentration in fetal bovine serum could be achieved by the standard addition method and without sample pre-treatment. The result showed a good agreement with the one determined by mass spectrometry, suggesting our biosensor as a promising tool for the detection of L-Arg in biological samples.