A Dual-Padded, Protrusion-Incorporated, Ring-Type Sensor for the Measurement of Food Mass and Intake.

Sensors (Basel, Switzerland) Hong W, Lee J, Lee WG
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Dietary monitoring is vital in healthcare because knowing food mass and intake (FMI) plays an essential role in revitalizing a person's health and physical condition. In this study, we report the development of a highly sensitive ring-type biosensor for the detection of FMI for dietary monitoring. To identify lightweight food on a spoon, we enhance the sensing system's sensitivity with three components: (1) a first-class lever mechanism, (2) a dual pad sensor, and (3) a force focusing structure using a ring surface having protrusions. As a result, we confirmed that, as the food arm's length increases, the force detected at the sensor is amplified by the first-class lever mechanism. Moreover, we obtained 1.88 and 1.71 times amplification using the dual pad sensor and the force focusing structure, respectively. Furthermore, the ring-type biosensor showed significant potential as a diagnostic indicator because the ring sensor signal was linearly proportional to the food mass delivered in a spoon, with R2 = 0.988, and an average F1 score of 0.973. Therefore, we believe that this approach is potentially beneficial for developing a dietary monitoring platform to support the prevention of obesity, which causes several adult diseases, and to keep the FMI data collection process automated in a quantitative, network-controlled manner.