2020

Short-Wave Infrared Fluorescence Chemical Sensor for Detection of Otitis Media.

ACS sensors Yim JJ, Singh SP, Xia A, Kashfi-Sadabad R, Tholen M, Huland DM, Zarabanda D, Cao Z, Solis-Pazmino P, Bogyo M, Valdez TA
阅读原文 PDF DOI PubMed

组成图示

示意图生成中

传感器类型

检测对象

检测原理

检测灵敏度

效应效果

传感器的构成

中文摘要

英文摘要

Otitis media (OM) or middle ear infection is one of the most common diseases in young children around the world. The diagnosis of OM is currently performed using an otoscope to detect middle ear fluid and inflammatory changes manifested in the tympanic membrane. However, conventional otoscopy cannot visualize across the tympanic membrane or sample middle ear fluid. This can lead to low diagnostic certainty and overdiagnoses of OM. To improve the diagnosis of OM, we have developed a short-wave infrared (SWIR) otoscope in combination with a protease-cleavable biosensor, 6QC-ICG, which can facilitate the detection of inflammatory proteases in the middle ear with an increase in contrast. 6QC-ICG is a fluorescently quenched probe, which is activated in the presence of cysteine cathepsin proteases that are up-regulated in inflammatory immune cells. Using a preclinical model and custom-built SWIR otomicroscope in this proof-of-concept study, we successfully demonstrated the feasibility of robustly distinguishing inflamed ears from controls (p = 0.0006). The inflamed ears showed an overall signal-to-background ratio of 2.0 with a mean fluorescence of 81 ± 17 AU, while the control ear exhibited a mean fluorescence of 41 ± 11 AU. We envision that these fluorescently quenched probes in conjunction with SWIR imaging tools have the potential to be used as an alternate/adjunct tool for objective diagnosis of OM.

关键词