This study reports on the development of a novel impedimetric immunosensor design using plant-derived antigenic glycoprotein for the detection of dengue virus (DENV) IgG antibodies. The electrochemical immunosensor platform was constructed using screen-printed carbon electrode (SPCE) modified with graphene/titanium dioxide (G/TiO2) nanocomposite to improve the electrode in terms electrochemical performance and specific surface area. A plant-derived dengue envelope domain III (EDIII) protein was used as the antigenic probe protein in this immunosensing strategy. Under optimised sensing conditions, the immunosensor demonstrated high sensitivity towards DENV IgG in a wide linear working range (62.5-2000 ng/mL), with a limit of detection of 2.81 ng/mL. The immunosensor showed high specificity for discriminating DENV IgG against antibodies of other infectious disease, including the closely related Zika virus (ZIKV). The reliability of the immunosensor in serological diagnosis was verified by challenging the immunosensor against serum samples, compared to conventional enzyme-linked immunosorbent assay (ELISA). As shown by its remarkable performance throughout the study, the devised immunosensor is proposed as a reliable and practical diagnostic tool for the serological detection of dengue in realistic applications.